WTW 1P21 CR 3200 Thermoreactors for Chemical Oxygen Demand, 2 x 12 Maximum Number of Samples , Same Program
Specification
Reactor type | Dry temperature control device with safety cover |
Cell shafts | 2 x 12 cell shafts for reaction cells, diameter: 16 ± 0.2 mm |
Reaction time setting | program settings or user-defined programs up to 180 min |
Temperature setting | stored program, user-defined settings from room temperature up to 170°C |
Controlling accuracy | ± 1 °C ± 1 digit |
Temperature stability | ± 0.5 K |
Overtemperature protection | 190 °C ± 5 °C |
Dimensions | D x W x H: 312 x 255 x 185 mm |
Weight: | 4 kg |
Output | unidirectional RS232 interface with 2 banana sockets for external temperature sensor (article-# 250 100), printer, PC |
Power supply | 115/230 VAC 50/60 Hz switchable via fuse setting allowed voltage tolerance: ±15 % maximum power consumption: 560 W, Fuse 2x6.3 AT |
Protection Class | IP 20 according to EN 60529 |
Ambient temperature | Operation +5 °C to +40 °C |
Climatic class | 2 according to VDI/VDE 3540 |
Relative humidity: Yearly mean: < 75 % | |
30 days /year: 95 % Other days: 85 % Light dew: yes | |
Site altitude: max. 5000 m above sea level | |
EMC | EN 61326-1 |
FCC 47 CFR Part 15 | |
Test certificates | cETLus, CE |
Safety standards | EN 61010-1 |
EN 61010-2-010 | |
UL 61010-1 | |
CAN/CSA C22.2#61010-1 | |
CAN/CSA C22.2#61010-2-010 |
- The CR 3200 has 2x12 cuvette slots, all of which can be opened with the same program. The CR 3200 also allows for eight custom programs with free selection of temperature up to 170 °C
Digestions of 24 round cells with pre-programmed and user-defined programs . In total 24 round cells (16 mm) in two seperate heating blocks using the same program
- 8 stored programs for easy and fast operation including COD fast digestion
- 8 user-defined programs from room temperature/25-170°C with high temperature digestion
- AQA/IQC with optional external temperature sensor for documentation
Programs for standard parameters COD including fast digestion (148°C, 20 min.), total nitrogen, total phosphorus, TOC and heavy metals of galvanizing with automatic heating and switch-off quick. User defined programs can be easily entered and stored from room temperature respectively 25-170°C to perform special digestions and thus allow high temperature digestion.
The very same program can be run for all vials at the same time; to perform two parallel digestion programs, the CR 4200 is the suitable choice. For the analytical quality assurance (AQA), the external temperature sensor TFK CR can be connected to the PC/printer instead of a round cell.
Programs for standard parameters
- 1: 148 °C for 120 minutes
- 2: 120 °C for 30 minutes
- 3: 120 °C for 60 minutes
- 4: 120 °C for 120 minutes
- 5: 100 °C for 60 minutes
- 6: 148 °C for 20 minutes
- 7: 150 °C for 120 minutes
- 8: 100 °C for 30 minutes