The Sovereign is a powerful workhorse, combining large capacity with clever design features designed to save you time and money in the running of your lab.
Key Features & Options
- 200 & 250 litre variants
- MP 25 Controller
- Ecomony round chamber
- Economical simple door closure
- Fixed seal
- Economical space saving design
- Drip tray under door
Option Requirements
- Air Supply at 38 psi, 7 cfm required for Air Ballasting option
- Post Cycle Drying will require AC701C vacuum option.
- Drain requirement of 150°C, 35Ømm copper pipe at floor level.
General Requirements
- † Sodium Chloride must not exceed a level greater than 500 ppm.
- † Use regular mains water only. DO NOT USE distilled or RO water
- DS Version requires a mains steam supply of 38psi 9kw 35lbs/hour
Complying Standards
- PED Cat 3 Module B+D Notified Body: British Engineering No. 0040
- Harmonised Standards applied: EMC BS EN 61326
- Community directives applied : Pressure Equipment Directive PED 2014/68/EU (formerly PED 97/23/EC)
- Designed in accordance with PD5500 (formerly BS5500) & BS2646
- Chamber and Door constructed of stainless steel BS-EN 10028-7:316
- ISO 9001: 2008
- CE Marked
Design Features
- 200 or 250 litres
- MP25 Controller
- Economy round chamber
- Robust space saving design
- Integral drip tray
- Grade 316 stainless steel chamber
- Anti door bolt over tight mechanism
- Complete versatility for almost any application
Optional Accessories
- †Loading Baskets
- †Waste Discard System
- †Fan Assisted High Speed Cooling
- †Condense Collector, which eliminates a drain require-
- †Air removal HEPA filter (for CAT3 environments)
- †Data Printers & Chart Recorders
- †Vacuum & Drying options
- †Anti Lime scale system
Capacities and dimensions (approx) |
||
Model |
Sovereign 200 |
Sovereign 250 |
Autoclave Weight |
470kg |
480kg |
Export Case Size |
990w x 1970h x 1610d |
990w x 1970h x 1610d |
Overall Dimensions |
830w x 1800h x 1450d |
830w x 1800h x 1450d |
Chamber Dimensions |
545 Ø x 860d |
545 Ø x 1057d |
Chamber Capacity |
200 Litres |
250 Litres |
Water Consumption (in chamber / typical per cycle) |
750ml |
750ml |
Water Consumption (Steam generator / typical per cycle) |
16 litres |
16 litres |
1L Media Bottles |
29 |
29 |
2L Media Bottles |
15 |
15 |
500ml Media Bottles |
82 |
82 |
NHS Discard Boxes |
H250xW320xL320 (3) |
H250xW320xL320 (3) |
Waste Discard Dimensions |
H198xW405xL470 (2) |
H198xW405xL470 (2) |
Trolley |
H1045xW580xL1290 |
H1045xW580xL1290 |
Safe Design Operating Limits |
||
Model |
Sovereign 200 |
Sovereign 250 |
Max working temperature |
138°C |
138°C |
Min working temperature |
7°C |
7°C |
Max ambient temperature |
28°C |
28°C |
Min ambient temperature |
11°C |
11°C |
Max hydraulic pressure |
4.14 Bar |
4.14 Bar |
Max working pressure |
2.62 Bar |
2.62 Bar |
Max working vacuum |
Minus 1 Bar |
Minus 1 Bar |
-
ManufacturerRodwell Autoclave Company
-
Country of OriginUnited Kingdom
-
Chamber Capacity (L)200, 250
-
Chamber Dimensions (mm)545 Ø x 860d, 545 Ø x 1057d
Standard Features
The common control system fitted to every autoclave in the Rodwell range. The MP25 has evolved to be one of the finest laboratory instrument control systems on the market today. What sets this system apart from any other control system is that it is totally interchangeable. The same MP25 control system can be used with any Rodwell autoclave, from a bench top to a large square chambered vacuum machine.
MP25 Options
25 programs
Can be programmed and stored by the laboratory Supervisor.
Smart card access system
- Allows programs to be stored and locked off by a supervisor
- Operator cards give access to pre-programmed cycles only
- Available with or without operator smartcard access for operator access
Load Activated Sterilise timer
(On/Off) prevents the sterilise timer from starting until the chamber and load have reached set point temperature.
Free steam timer (adjustable by time)
Holds the chamber at approximately 100°C (without pressurising) for a pre-determined time during the heating period and allow efective steam penetration into the load.
Delayed start (24 hour clock)
Timer option allows the autoclave to be pre-programmed at a pre-determined time so the autoclave completes a cycle just before the lab arrives for work the following morning for pre-prepared warm media.
Adjustable Stay Warm (on/off)
Keeps the chamber warm at the end of the cycle for pourable media. This is adjustable to allow the operator to select the optimum temperature.
Instrument Cycles (programs 21 to 24 only)
Exhausts the chamber at the end of the cycle for dryer glassware and instruments upon completion.