The Grant bio UVC/T-M-AR is an efficient general-purpose stainless steel PCR UV cabinet workstation for carrying out short-term procedures.
The UVC/T-M-AR cabinets are equipped with an open UV lamp installed in the upper part of the hood. UV-radiation from these lamps effectively disinfects the working area chamber, inactivating DNA / RNA fragments during 15-30 minutes of exposure. The digital timer controls the duration of the direct UV irradiation. An additional daylight lamp provides a bright, clear working area. Models are equipped with a high efficiency flow-type bactericidal UV cleaner-recirculator which provide constant decontamination inside the cabinet during operation. The UV cleaner-recirculator consists of UV lamp, dust filters and a fan. The recirculation effect increases the maximum density of the UV light making it sufficiently effective for DNA / RNA inactivation. The recirculator processes 360 UV-cabinet volumes per hour, creating permanently aseptic conditions within the cabinet. Designed for robustness and ease of access, it features a built-in UV cleaner-recirculator to prevent unwanted contamination and to protect the user from direct UV light during manipulation.
Applications These PCR UV-cabinets are designed for clean operations with DNA samples and provide protection against contamination. They are not recommended for operations with dangerous infectious and viral materials as well as DNA / RNA amplification. |
The Grant bio UVT-B-AR is an economical general purpose PCR UV cabinet in for carrying out short-term procedures. The UVT-B-AR cabinets are equipped with an open UV lamp installed in the upper part of the hood. UV-radiation from these lamps effectively disinfects the working area chamber, inactivating DNA / RNA fragments during 15-30 minutes of exposure. The UV light switches off automatically when door is opened. The model has a stainless steel work surface, metal framework and shock proof glass panels. The built-in UV cleaner-recirculator both increases the maximum density of UV light and generates 25 m³/h air flow exchange - prevents unwanted contamination and protects the user from direct UV light during manipulation. The UV cleaner-recirculator consists of UV lamp, dust filters and a fan. The recirculation effect increases the maximum density of the UV light making it sufficiently effective for DNA / RNA inactivation. The recirculator processes 360 UV-cabinet volumes per hour, creating permanently aseptic conditions within the cabinet. The digital timer controls the duration of the direct UV irradiation. An additional daylight lamp provides a bright, clear working area.
Applications Life science - germicidal and virucidal, inhibition of DNA and RNA contamination, applications requiring no residual decontaminants such as disinfectants, operations with DNA/RNA amplicons, microbial research. |
The Grant bio UVT-S-AR is a large capacity stainless steel PCR UV cabinet workstation with additional space for equipment and accessories to allow for more comfortable and convenient working in PCR applications. These UVT-S-AR cabinets are equipped with an open UV lamp installed in the upper part of the hood. UV-radiation from these lamps effectively disinfects the working area chamber, inactivating DNA / RNA fragments during 15-30 minutes of exposure. The digital timer controls the duration of the direct UV irradiation. An additional daylight lamp provides a bright, clear working area. They are equipped with a high efficiency flow-type bactericidal UV cleaner-recirculator which provide constant decontamination inside the cabinet during operation. The UV cleaner-recirculator consists of UV lamp, dust filters and a fan. The recirculation effect increases the maximum density of the UV light making it sufficiently effective for DNA / RNA inactivation. The recirculator processes 360 UV-cabinet volumes per hour, creating permanently aseptic conditions within the cabinet.
Applications These PCR UV-cabinets are designed for clean operations with DNA samples and provide protection against contamination. They are not recommended for operations with dangerous infectious and viral materials as well as DNA / RNA amplification. |
Specifications
Dimensions |
h x d x w mm |
|
||||||
UVC/T-M-AR |
UVT-B-AR |
UVT-S-AR |
||||||
General purpose |
General purpose economy |
PCR workstation |
||||||
UVC/T-M-AR |
UVC/T-M-AR SKT |
UVT-B-AR |
UVT-B-AR INL |
UVT-S-AR |
||||
535 x 515 x 690 |
555 x 585 x 690 |
585 x 585 x 1245 |
||||||
Construction |
Stainless steel frame and working area |
|||||||
Panels |
Glass with UV-protective film |
|||||||
Front opening with three adjustable positions |
• |
|||||||
Open UV lamp, 25W bactericidal, 254nm, ozone free |
1 |
- |
||||||
Open UV lamp, 30W bactericidal, ozone free |
- |
2 |
||||||
Bactericidal air recirculator, 25m3/h air flow exchange |
• |
|||||||
UV recirculator, 25W (efficiency >99% per 1 cycle) |
1 |
- |
||||||
UV recirculator, 30W (efficiency >99% per 1 cycle) |
- |
1 |
||||||
White lamp for workplace illumination |
15W |
1 |
- |
|||||
30W |
- |
1 |
||||||
Radiation type |
Ultraviolet (253.7m), ozone free |
|||||||
Optical transmission |
95% |
|||||||
Digital timer |
0 to 24 hours |
• |
||||||
Internal power outlets |
- |
1 |
- |
3 |
||||
Internal working area |
mm |
645 x 490 |
1210 x 500 |
|||||
Flow rate |
m3/h |
7 |
||||||
Power consumption |
230V W |
67 (0.3A) |
530 (4.5A) |
|||||
120V W |
- |
315 (1.4A) |
||||||
Nominal operating voltage V |
120 or 230 (50-60Hz) |
120 (60Hz) or 230 (50Hz) |
||||||
Weight |
kg |
31 |
31 |
58 |