- Digital models enable highly accurate temperature precision
- Fast heat-up rate: 25°C to 100°C in 7-9 minutes, depending on the model chosen
- Advanced temperature control. Temperature stability at 37°C ± 0.1°C
- Temperature range from from ambient +5°C to 100°C or 200°C, depending on the model chosen
- Choice of models with 2, 3 or 4 interchangeable blocks
- Versatile range of interchangeable heating blocks to fit all popular sample tubes and plates
Electrical Safety
The Heating Block range has been designed to comply with all RF interference and electrical safety regulations and are CE marked.
Warranty
Cole-Parmer® Heating Block have earned a strong reputation for reliability. All instruments are manufactured in the UK and supplied with a 3 year warranty. The anodised aluminium Heating Block inserts are supplied with a 1 year warranty.
Applications
Applications include incubation, boiling, inactivation, wet washing, sample concentration, enzyme digestions, enzyme activity studies, nucleic acid hybridisations and many other clinical and industrial purposes. Larger three and four-block models are available for high throughput applications such as Chemical Oxygen Demand. The innovative Cole-Parmer® range also includes a twin control unit, accommodating two blocks with independent digital
temperature controls. Each block can be set to different temperatures - ideal for multiple users or for applications where samples have to be transferred between two temperatures very quickly.
Countdown Timer
Each digital unit features a countdown timer, which can be adjusted to time your experiment from anywhere between 1 minute and 100 hours, in 1 minute increments. To set the timer, press and hold the timer button and press the up or down buttons. Each press of the up or down buttons will increase or decrease the set time by 1 minute. If the buttons are held down, the time change will accelerate to 200 min per second. Release the timer button to activate the timer. An audible beep will indicate the timer is active. When the count-down has completed, an audible beep will be heard and the display will alternate between END and current temperature. Press the set/reset button to cancel the beeper. The unit will cool to ambient once the count-down time has completed. To cancel the timer, press and hold the set/reset button until STOP appears and an audible beep is heard.
Over-temperature cut-out
In the event of a fault occurring allowing the Heating Block to overheat, a thermal fuse will cut in and remove power from the heater when the temperature reaches approximately 192°C in BH250-2, BH250-3 models and 240°C in BH250D models - eliminating any risk.
Accurate temperature control
Highly accurate digital temperature control is made quick and easy with a bright 5-digit orange LED display. Each unit has a set point resolution of 0.1°C with temperature range from ambient + 5°C up to 100°C or 200°C depending on the model chosen. The Cole-Parmer® Heating Block facilitate a fast heat-up rate from 25°C to 100°C in 7-9 minutes, depending on the model chosen and all models are able to achieve a temperature stability at 37°C ± 0.1°C.
New panel venting
Each Cole-Parmer® digital Heating Block has venting to the side panels. Whilst helping to create a fresh modern look, the vents are designed to dissipate heat from the outer surfaces of the Dri-Block®, allowing users to safely handle and relocate their Heating Block even after extended use at 100˚C or 200˚C
Technical Specifications
Specification |
BH-250D-2 |
BH-250D-2TC |
BH-250D-3 |
BH-250D-4 |
BH-250D-2-HT |
BH-250D-3-HT - BH-250D-3-HT-RS |
Blocks |
2 |
2 |
3 |
4 |
2 |
3 |
Display |
5 character LED |
5 character LED |
5 character LED |
5 character LED |
5 character LED |
5 character LED |
Working range |
Ambient +5°C to 100°C |
Ambient +5°C to 100°C |
Ambient +5°C to 100°C |
Ambient +5°C to 100°C |
From Ambient +5°C to 200°C |
From Ambient +5°C to 200°C |
Minimum set temperature |
0°C |
0°C |
0°C |
0°C |
0°C |
0°C |
Resolution |
0.1°C |
0.1°C |
0.1°C |
0.1°C |
0.1°C |
0.1°C |
Accuracy (37°C) |
< ±1°C |
< ±1°C |
< ±1°C |
< ±1°C |
< ±1°C |
< ±1°C |
Accuracy (100°C) |
< ±1°C |
< ±1°C |
< ±1°C |
< ±1°C |
< ±1°C |
< ±1°C |
Stability (37°C) |
< ±1°C |
< ±1°C |
< ±1°C |
< ±1°C |
< ±1°C |
< ±1°C |
Stability (100°C) |
< ±0.15°C |
< ±0.15°C |
< ±0.15°C |
< ±0.15°C |
< ±0.15°C |
< ±0.15°C |
Max variation within block (37°C) |
< 0.2°C |
< 0.2°C |
< 0.2°C |
< 0.2°C |
< 0.2°C |
< 0.2°C |
Max variation within block (100°C) |
< 0.25°C |
< 0.25°C |
< 0.25°C |
< 0.25°C |
< 0.25°C |
< 0.25°C |
Max variation between blocks (37°C) |
< 0.5°C |
< 0.5°C |
< 0.5°C |
< 0.5°C |
< 0.5°C |
< 0.5°C |
Max variation between blocks (100°C) |
< 1.0°C |
< 1.0°C |
< 1.0°C |
< 1.0°C |
< 1.0°C |
< 1.0°C |
Heat up time 25-100°C (min) |
8 |
7 |
7 |
7 |
8 |
9 |
Heat up time 25-200°C (min) |
na |
na |
na |
na |
23 |
19 |
Audible alarm |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
Minimum timer |
1min |
1min |
1min |
1min |
1min |
1min |
Maximum timer |
99hr 59min |
99hr 59min |
99hr 59min |
99hr 59min |
99hr 59min |
99hr 59min |
Safety cutout device |
Thermal fuse |
Thermal fuse |
Thermal fuse |
Thermal fuse |
Thermal fuse |
Thermal fuse |
Dimensions (WDH) |
356 x 260 x 105mm |
356 x 260 x 105mm |
356 x 260 x 105mm |
356 x 260 x 105mm |
356 x 260 x 105mm |
356 x 260 x 105mm |
Electrical supply |
230V or 115V 50-60Hz |
230V or 115V 50-60Hz |
230V or 115V 50-60Hz |
230V or 115V 50-60Hz |
230V or 115V 50-60Hz |
230V or 115V 50-60Hz |
Power (W) |
300 |
450 |
450 |
600 |
300 |
450 |
Shipping weight (Kg) |
5 |
6 |
6 |
7 |
5 |
6 |
Warranty (Years) |
3 |
3 |
3 |
3 |
3 |
3 |
-
ManufacturerAntylia Scientific
-
Country of OriginUnited Kingdom
-
Number of Blocks2, 3, 4
-
Working RangeAmbient +5°C to 100°C, Ambient +5°C to 200°C
-
Voltage230V, 115V
-
Model NoBH-250D-2, BH-250D-2-115, BH-250DC-2, BH-250DC-2-115, BH-250DC-3, BH-250DC-3-115, BH-250DC-4, BH-250DC-4-115, BH-250D-2-HT, BH-250D-2-HT-115, BH-250D-3-HT, BH-250D-3-HT-115, BH-250D-3-HT-RS, BH-250D-3-HT-RS-115
-
LegacyDB100/2, DB100/2/115, DB100/2TC, DB100/2TC/115, DB100/3, DB100/3/115, DB100/4, DB100/4/115, DB200/2, DB200/2/115, DB200/3, DB200/3/115, DB200/3/RS, DB200/3/RS/115