WTW 1P22- CR 3200 Thermoreactor Digestions of 24 Round Cells

Not rated yet
Reactor typeDry temperature control device with safety cover
Cell shafts2 x 12 cell shafts for reaction cells, diameter: 16 ± 0.2 mm
Reaction time settingprogram settings or user-defined programs up to 180 min
Temperature settingstored program, user-defined settings from room temperature up to 170°C
Controlling accuracy± 1 °C ± 1 digit
Temperature stability± 0.5 K
Overtemperature protection190 °C ± 5 °C
DimensionsD x W x H: 312 x 255 x 185 mm
Weight:4 kg
Outputunidirectional RS232 interface with 2 banana sockets for external temperature sensor (article-# 250 100), printer, PC
Power supply115/230 VAC 50/60 Hz switchable via fuse setting
allowed voltage tolerance: ±15 % maximum power consumption: 560 W, Fuse 2x6.3 AT
Protection ClassIP 20 according to EN 60529
Ambient temperatureOperation +5 °C to +40 °C
Climatic class2 according to VDI/VDE 3540
 Relative humidity: Yearly mean:
 30 days /year: 95 % Other days: 85 % Light dew: yes
 Site altitude: max. 5000 m above sea level
EMCEN 61326-1
 FCC 47 CFR Part 15
Test certificatescETLus, CE
Safety standardsEN 61010-1
 EN 61010-2-010
 UL 61010-1
 CAN/CSA C22.2#61010-1
 CAN/CSA C22.2#61010-2-010
Price Upon Application Or Discontinued

Digestions of 24 round cells with pre-programmed and user-defined programs

In total 24 round cells (16 mm) in two seperate heating blocks using the same program

• 8 stored programs for easy and fast operation including COD fast digestion
• 8 user-defined programs from room temperature/25-170°C with high temperature digestion
• AQA/IQC with optional external temperature sensor for documentation

Programs for standard parameters COD including fast digestion (148°C, 20 min.), total nitrogen, total phosphorus, TOC and heavy metals of galvanizing with automatic heating and switch-off quick. User defined programs can be easily entered and stored from room temperature respectively 25-170°C to perform special digestions and thus allow high temperature digestion.

The very same program can be run for all vials at the same time; to perform two parallel digestion programs, the CR 4200 is the suitable choice. For the analytical quality assurance (AQA), the external temperature sensor TFK CR can be connected to the PC/printer instead of a round cell.

Programs for standard parameters:

1: 148 °C for 120 minutes
2: 120 °C for 30 minutes
3: 120 °C for 60 minutes
4: 120 °C for 120 minutes
5: 100 °C for 60 minutes
6: 148 °C for 20 minutes
7: 150 °C for 120 minutes
8: 100 °C for 30 minutes

    • SKU
      1P22-
  • Manufacturer
    WTW - Xylem Inc