Designed to save space on your laboratory workbench, the 2840EL-D is a front loading benchtop autoclave that provides high quality, repeatable performance and accountability for a wide range of applications used in modern laboratories.
The 2840EL-D has an advanced multi-color control panel and a chamber made of 316L stainless steel. Each model has a number of optional added-value features which can be configured for fast cooling, efficient drying, biohazard and waste sterilization, Fo control, and more.
Data & Software
The 2840 EL gives you easy access to its cycle data. Via USB, Ethernet or printer.
- Print a hard copy of cycle data via optional printer
- Store 200 cycles on your autoclave for reprinting or downloading to a PC
- Connect to a local network for use with the optional RPCR Software for remote monitoring of up to 8 autoclaves
- Generate PDF reports with graphs and tables using optional RPCR Software
- Multilevel Password protection
Technical Specifications
- Overall Dimensions (WxHxD) - 530mm x 440mm x 630mm
- Chamber Dimensions Ø x D - 280mm x 400mm
- Chamber Volume - 28 Litres
- Sterilization Temp. Range - 105°C to 138°C
- Voltage & Frequency 1 Phase 208V, 230V 50-60 Hz
Sterilization Programs
30 program cycles are available with each autoclave. Up to 8 cycle programs are factory set according to optional features. The remaining cycles are fully customizable by the user.
Program Cycles and Features
- Solid and glassware loads at 134°C or 121°C for delicate loads (plastics)
- Liquid loads and waste liquids at 121°C
|
Description |
Standard Cycles |
- Solid and glassware loads at 134°C or 121°C for delicate loads (plastics)
- Liquid loads and waste liquids at 121°C
|
Standard Cycles added with optional features |
- Cooling Coil + Compressed Air
- Fast cooling for liquid loads at 121°C
|
Feature: Biohazard Air Filter |
- During air removal all exhaust air is filtered through a 0.2 μm biological filter to prevent contamination of the laboratory
- Biohazard solid loads at 134°C
- Biohazard liquid loads at 121°C
|
Feature: Vacuum Pump |
- Pre-vacuum cycles
- Solid and glassware loads at 134°C or 121°C for delicate loads (plastics)
- Liquid loads and waste liquids at 121°C
- Air Leakage Test Cycle
|
Feature: Vacuum Pump + Steam Generator |
- Pre and Post vacuum cycles
- Hollow, porous and textile loads at 134°C
- Waste: hollow, porous and textile at 121°C
- Bowie & Dick steam penetration test at 134°C
|
Media Processing Cycles |
- Isothermal Processing: for preparing agar and other biological media with a temperature range from 60°C to 95°C that allows for gentle heating and cooling down of agar
- Holding Temperature: Special program with programmable holding temperature at the end of the cycle to prevent cooling of media
|
Special Custom Cycles |
- Tuttnauer is able to provide specifically customized cycles upon request. These may include material stress test, ageing test, varnish test, and others.
- Extended Sterilization Time: Special program with extended sterilization times up to 999 minutes
- Multiple Cycle (Material Sterss Test): Special program to automatically run multiple cycles on the same load
|