Class B Efficiency with Maximum Reliability
Designed to Meet Your Healthcare Clinics’ Everyday Sterilization Needs
T-Top, a 22 litre tabletop autoclave, specifically developed for dental and medical clinics of small to medium practices. Simplicity and economy, this tabletop autoclave covers the fundamental needs for your general clinic sterilization with the aim of increasing the productivity of your practice.
T-Top offers universal capacity, handed to you, in an affordable and reliable solution you can count on for managing your daily practice sterilization process workload.
The perfect balance between performance, ergonomics, technology & usability. Resulting in a strong and efficient autoclave which perfectly meets the requirements of all standard practices.
T-Top is a sterilizer with high performance class B cycles, user friendly advanced technology with Wi-Fi connectivity, and an intuitive colour touchscreen.
- 22 litre Chamber
- Plug and Play. No dedicated installation connections needed, simply a power source
- Reliable Performance. Quality product that will do all you need and expect
- Traceability & Connectivity. Connect using Ethernet connection, USB ports, Built-in Wi-Fi, and R.PC.R
- Time saver setting. Set test cycles and sterilization cycles to start at any time
- Compact Design. Does not require a lot of counter top or shelf space
- Meets strictest international standards
Optimized Cycle Programs For Efficient And Consistent Performance
Program name | Sterilization time | Load type | Type of use |
---|---|---|---|
Unwrapped 134°C | 4 min. | Solid / porous load | For immediate use |
Wrapped 134°C | 4 min. | Solid / porous load | For storage |
Unwrapped 121°C | 20 min. | Solid / porous load | For immediate use |
Wrapped 121°C | 20 min. | Solid / porous load | For storage |
Prion 134°C | 18 min. | Solid / porous load | For storage |
- Bowie & Dick Test Cycle – Chemical indicator in a product challenge device. Verify effective steam penetration. Also suitable for testing air removal with Helix test.
- Leak Test Cycle – Verify the integrity of the pressure vessel and its plumbing on a periodic basis as referred to in ISO 17665-1.
Chamber size
Volume
Overall Dimensions (D x W x H)
Usable chamber space
Overall dimensions
Max # of trays
Tray dimensions
Weight
Digital Connections
Water filling
Water Tanks
|
T-Top Meets the Following International Standards
CE MDR
Certificate number 3903015CE01
Valid until 01-03-2028
Electromagnetic compatibility
IEC 61326-1
Electrical safety
IEC 61010-1, 61010-2-040
Verification & validation functional tests
EN 13060
Software validation
EN 62304, FDA guidance
Max # of trays
5 Trays
Quality Management System
ISO 9001:2015
ISO 13485:2016
Pressure Vessel
PED 2014/68/EU