The AD-1690 instantly identifies any compromise in the pressure integrity of small pumps, piping, storage containers and especially micropipettes. Tests devises to 50ml capacity.
- Quick testing in less than 5 seconds
- Leak detection by depressurization ensuring no dust enters the object being measured
- Replaceable air filter to protect the AD-1690 from dust when it intakes air
- RS-232C interface to output data to a PC or an A&D peripheral device
- Three types of attachments to cover micropipettes with capacities ranging from 2 μL to 10,000 μL
Specifications
Depressurization value | -20 kPa ± 20% (Fixed) |
Pressure tolerance inside instrument | +0.2 kPa / 10 seconds |
Leak test conditions | Pressure tolerance setting: +0.1 to +20 kPa (Unit: 0.1kPa) Monitoring time: From 0.5 seconds (Unit: 1 second) |
Pump operation time | Up to 6 seconds |
Dimensions | (W) 231 mm x (D) 126 mm x (H) 78 mm |
Weight of the main unit | Approx. 570 g |