Specifications - pH Electrode InLab Max Pro-ISM
Type of sensor | Combined pH electrode with temperature probe and ISM |
Measuring range | pH 0 – 14 |
Temperature range | 0 °C – 100 °C |
Cable and connectors | MultiPin™ |
Type of junction | Immovable glass sleeve |
Reference System | ARGENTHAL™ with Ag+-trap |
Reference electrolyte | FRISCOLYT-B® |
Type of membrane glass | HA glass |
Membrane resistance (25°C) | < 600 MΩ |
Shaft Material | Glass |
Shaft Diameter | 12 mm |
Shaft length | 120 mm |
ISM | Yes |
Temperature probe | NTC 30 kΩ |
Base Unit Temperature Range (Metric) | 0 °C – 100 °C |
A glass-body, combination pH electrode with integrated temperature sensor, MultiPinTM head and ISM. Designed for general laboratory applications; highly reproducible and fast.
Self-cleaning junction
The high electrolyte outflow prevents the pollutants from entering and clogging the junction, which would interfere with the reading.
High pH values, low alkali error
The extremely robust HA membrane glass is specifically designed for use at high pH values, reducing otherwise common alkali errors to a minimum.
ISM® - Intelligent Sensor Management
All the important information is stored in the sensor and transferred automatically to the meter. The ideal support for fast and safe results.
ISM® - Intelligent Sensor Management models store calibration data and automatically transfer the data to your meter
Choose open junction versions for dirty and complex samples
Cables not included with all electrodes for additional versatility—select one that connects to your meter
ISM technology stores calibration data in the sensor and automatically transfer it to the meter.⌡Select the appropriate sensor cable for your application, meter, and ATC connection—BNC, waterproof BNC, DIN, and Lemo OO cables available with banana or RCA (cinch) connectors.⌡Sleeve junction provides fast results and works best in dirty samples. High alkali glass membrane provides low-alkali error—ideal for high pH and high temperature applications. ARGENTHAL™ reference system has a silver ion trap, for silver-ion-free electrolyte. No clogging junction from sulfides or protein binding—perfect for biological samples and TRIS buffers.