LEEC GA2000 Research CO2‚ Incubators, Fan Assisted with 95 °C Decontamination Facility, 150 Litre Capacity, 600W

ModelCapacityExternal - mmInternal - mmShelvesWeightPower Rating
GA2000150 litres880h x 635w x 660d600h x 510w x 500d


4


85kg


600W


SPECIFICATION

Model


Research models

(GA2000 / GA3000)

TEMPERATURE RANGE

 CONTROL

VARIATION

RECOVERY

 

5° above ambient to +60°C

<±0.1°C at 37.0°C

<±0.2°C at 37.0°C

Typically <6 minutes

 

CO2 RANGE

CONTROL

RECOVERY

CO2 DETECTOR

0.1 to 20.0%

<±0.2° at 5.0%

Typically 3 minutes

Infrared (drift free)



RH RANGE

 

All models: Ambient or 95% by forced evaporation

 

CONSTRUCTION

 

Stainless steel chamber (304 grade). Polyester powder coated stainless steel body with anti-microbial paint as standard. Heated inner glass door(s). Fan assisted. CFC free insulation.

 

ALARMS

High / Low Temperature

High / Low CO2

Time Delay

Remote Connections

HIGH TEMPERATURE DECONTAMINATION FACILITY (95°C)YES
  • £4423.00
Qty:  

LEEC Research and standard CO₂ incubators set advanced standards in performance and freedom from contamination. The two Research models have the added benefit of a high temperature decontamination facility (95°C).

Features

  • Advanced microprocessor control system
  • Infrared CO2 detector
  • Soft touch control panel
  • Bright LED displays for temperature and CO2
  • Comprehensive alarm system
  • Built in cooling coil for use with an external LEEC cooler unit
  • 95°C decontamination facility (GA2000 and GA3000)
  • CO2 HEPA filter
  • Heated inner door(s)
  • 150 or 320 litre chamber capacities
  • Stainless steel shelves

 

ADVANCED MICROPROCESSOR CONTROL

The specially developed microprocessor controller gives the user day to day control over all alarm settings, alarm delay and calibration adjustment. These settings are protected by an anti-tampering procedure. Sensors placed within key areas of the chamber constantly monitor and enable the control system to optimise temperature and CO2 levels. After door opening, the culture conditions quickly recover without overshooting, and stable control is maintained. Profiled heating, a technique developed by LEEC, eliminates all risk of "hot-spots".

LEEC GA CO2 incubators (regardless of age) can become cooled CO₂ incubators when connected to a LEEC self contained recirculating cooler unit.

ACTIVE CONTAMINATION CONTROL

All models have fan assisted uni-directional airflow in the chamber, an in-line HEPA filter and an ultra smooth stainless steel chamber with rounded corners. Indirectly heated inner glass door(s) remain condensation free. These features all help to reduce the risk of contamination. The two Research models (GA2000 and GA3000) also have a high temperature decontamination facility (95°C) which is operated by a key. User experience has shown high temperature to be especially effective in eliminating culture contamination.

IN-VIVO SIMULATION

The LEEC designed multi-function microprocessor control system combined with fan assisted airflow ensures that temperature and CO2 stabilities (typically <±0.2°C and <±0.2%) are maintained to the tightest tolerances in order to give optimum "in-vivo" simulation.

 

 

STANDARD FEATURES

CONTROL:-

  • Infrared CO2 sensor
  • Microprocessor control with soft touch control panel.
  • Bright LED displays for temperature and CO2.
  • Fan assisted chamber for precise culturing conditions.

 

CONSTRUCTION:-

  • Steel outer cabinet for longer life.
  • Ultra smooth stainless steel chamber.
  • Stainless steel perforated shelves.
  • Adjustable front levelling feet.
  • Heated inner glass door(s).
  • Two 12mm cable access ports for monitoring probes.
  • Fan assisted chamber with uni-directional airflow.
  • White polyester powder coat paint.

 

HEATING AND SAFETY:-

  • 3 sided heating system.
  • Fan assisted chamber with uni-directional airflow.
  • Heated inner door(s).
  • High and low alarms (temperature)
  • High and low alarms (CO2)
  • Independent over temperature safety cut out
  • Remote alarm connections (volt-free, N/O, N/C)
  • Built-in copper cooling coil (can be connected to a LEEC Self Contained Recirculating Cooler Unit for chamber temperatures near or below ambient temperature).

 

CONTAMINATION CONTROL:-

  • HEPA filter removes contaminants from CO2 gas, prior to injecting into the chamber.
  • Key operated high temperature decontamination system 95°C (models GA2000 and GA3000 only).
  • White polyester powder coat paint with anti-microbial (silver based) additive. 

 

Every LEEC Incubator is made to order to meet the exacting needs of every customer.  LEEC prides itself on being able to deliver exactly what the customer wants and relishes the challenge of producing a Incubator that is tailored to every specific customer need.

    • SKU
      GA2000
  • Gross Capacity (L)
    150
  • Dimensions HxWxD (mm)
    880 x 635 x 660
  • Temperature Range (°C)
    5° above ambient to +60°C
  • Construction
    Stainless Steel Interior, Enamelled Steel Exterior
  • CO2 range [Vol.-% CO2]
    0.1 to 20
  • Humidity range [% RH]
    Ambient or 95% by forced evaporation
  • Manufacturer
    LEEC Limited