Hamilton ML615-DIL Microlab 615 Basic Dual Syringe Diluter, 100-240V 1.5A max 50-60 Hz

Not rated yet
0

Hamilton Microlab 600 Series Dual Syringe Diluter with Basic Controller
The ML615-DIL ships complete with the Concorde Hand Probe, Universal Valves, Fill/Dispense Tubing, Accessory Holder, country specific Power Cord, and the choice of two syringes 
If no syringes are selected at the time of the order, a 2.5 mL and 250µl syringe will be included.


Bubble Free Prime syringe
For any syringe pump, the key to achieving the most accurate dispenses is eliminating all air from the fluid path
Traditional syringes trap approximately 50µl of air between the tip of the syringe and the valve, which makes small syringes difficult to prime
The Bubble Free Prime (BFP) syringe has a conical plunger tip that extends through the threaded termination and into the valve 
This unique design expels the air from the syringe and valve decreasing the number of priming cycles required.

Universal Valves
The high precision Universal Valves of the ML600 are robust and flexible
The innovative fluid logic enables the same Universal Valve to be used in all MICROLAB 600 dispensing and diluting applications
Simply interchange the valve plugs and tubing to achieve the required configurations.

Hamilton ML615-DIL Microlab 615 Basic Dual Syringe Diluter, 100-240V 1.5A max 50-60 Hz Specifications

Power Rating 24 VDC, 2.5A
Power Requirements 100-240V 1.5A max 50-60 Hz
Volume Left Syringe Custom Volume Left Syringe
Volume Right Syringe Custom Volume Right Syringe
Program Memory One program stored in memory while power is on
Syringe Drive Mechanism Stepper motor driven high precision lead screw with encoder




Call 01463 830401 For Any Assistance

The Microlab 600 Series Diluter 

The Microlab 600 is a highly precise syringe pump with a graphical user interface designed to quickly and easily dilute and dispense liquids. This positive displacement system provides better than 99% accuracy, independent of a liquid's viscosity, vapor pressure and temperature. The inert fluid path minimizes sample carry over and improves compatibility with harsh chemicals.

Where is the Microlab 600 used?

Every laboratory has tasks that are too small to automate and too large to reliably accomplish by hand. The Microlab 600 is a semi-automated liquid handler designed specifically for these in-between applications. The Microlab 600 increases throughput and consistency while reducing cost and wasted buffer. Common industries utilizing the Microlab 600 for sample and standard preparations include forensics, environmental analysis, mining, manufacturing and many more.

PTFE Tube vs. Disposable Tip

For most applications the PTFE Tube Diluter is the preferred choice because positive displacement ensures expectational accuracy without technique variability. The unique design flushes the tubing between samples so carry-over is not an issue even for applications like trace metals analysis. However, some applications like PCR are more sensitive to carry-over while other applications like forensics have more regulatory scrutiny making disposable tips the best option.

 

The Controller

The Microlab 600 Controller integrates a streamlined user interface with a large touch screen display. Dilutions, dispenses, titrations and more are visually displayed in real-time with just the touch of a button.

A Hardware Key unlocks the Advanced functionality inluding Wizards and Custom Methods. The Hardware Key also provides an upgrade path for the syringe pump and controller to the most current firmware.

 

Specifications Microlab 600 Controller
Screen Size 5.7 inch (15cm diagonal)
Screen Resolution 640 (H) x 480 (V) pixels
Tilt Positions 5 positions from 90° to flat
Mounting Options On top of the Syringe Pump, Bench Top, or Wall Mount
Program Memory 2 GB (With Advanced Upgrade)
Communication Type Ethernet, 10/100 BASE-T
Power Rating 24VDC, 2.5A
Dimensions 2.25 x 6.5 x 7 inch
(57.2 x 165.1 x 177.8mm)
in down position
Weight 1.9 lbs (0.86 kg)

 

 

The Syringe Pump

The Microlab 600 is available as a single or dual syringe system. The high torque, precision stepper motors provide unsurpassed positional accuracy across the full range of Hamilton syringes from 10 µL to 50 mL.

Specifications Microlab 600 Syringe Pump
Accuracy +/- 1%
Precision +/- 0.2%
Flow Rate 0.003 - 6000 µL/second (depending on the syringe that is selected)
Syringe Resolution 0.002% of the nominal syringe volume
Compatible Syringes 10, 25, 50, 100, 250, 500 µL, 1, 2.5, 5, 10, 25, and 50 mL
Fluid Path Borosilicate, PTFE, CTFE
Communication Type Ethernet, 10/100 BASE-T
Communication Protocol .NET 2.0 Application Programming Interface (API)
Pump Memory One method stored in non volatile memory
Calibration Factory tested and traceable to N.I.S.T. standards
Certifications CE, CSA
Power Rating 24 VDC, 2.5A
Dimensions 7 x 5.5 x 10.5 inch (177.8 x 139.7 x 266.7 mm)
Weight 13 lbs (5.9 kg)

 

 

General Microlab 600 Accuracy & Precision

Syringe Size (µL) Percent Stroke Accuracy (± %) Precision (%)
10 5% ≤ Stroke < 30% 3.0 2.0
30% ≤ Stroke 2.0 0.2
25 5% ≤ Stroke < 30% 3.0 2.0
Stroke = 30% 1.5 0.2
Stroke = 100% 1.0 0.2
50 5% ≤ Stroke < 30% 3.0 1.5
Stroke = 30% 1.2 0.5
Stroke = 100% 1.0 0.2
100 5% ≤ Stroke < 30% 3.0 1.5
Stroke = 30% 1.2 0.5
Stroke = 100% 1.0 0.2
250 5% ≤ Stroke < 30% 3.0 1.5
Stroke = 30% 1.2 0.5
Stroke = 100% 1.0 0.2
500 1% ≤ Stroke < 5% 3.0 1.5
5% ≤ Stroke < 30% 1.2 0.5
Stroke = 30% 1.0 0.2
Stroke = 100% 1.0 0.2
1,000 & larger 1% ≤ Stroke < 5% 3.0 1.5
5% ≤ Stroke < 30% 1.2 0.5
Stroke = 30% 1.0 0.2
Stroke = 100% 1.0 0.2
    • SKU
      ML615-DIL
  • Manufacturer
    Hamilton Company