FRYKA mini-freezers are the ideal solution when small samples directly cooled in the workplace should be available. With content of 8 and 18 litres they can be udes in all scientific and
industrial laboratories for cooling and freezing.
Technical Data
Model | KBT 08-51 | KBT 08-51 + ST100 | KBT 18-41 U ST100 |
Temperature range [min/max] | -50° c to -30° C | -50° c to -30° C | -40° c to 0° C |
Control accuracy | +/- 1 K | +/- 1 K | +/- 1 K |
Control | ST71 | ST71 | ST100 |
Volume (Litres) | 8 | 8 | 18 |
Circulation air fan | Yes | ||
Interior [WxDxH] | 15 x 30 x 20 cm | 15 x 30 x 20 cm | 30 x 27 [32) x 20 cm |
External dimensions [WxDxH] (cm) | 36 x 49 x 57 | 36 x 49 x 57 | 52 X 52 X 58 |
Weight (kg) | 36 | 36 | 50 |
Electrical Connection | 230 V / 50 Hz | 230 V / 50 Hz | 230 V / 50 Hz |
Ambient temperature [min/max] |
+ 12° C / +30° C | + 12° C / +30° C | + 12° C / +30° C |
|
|
|
Characteristics:
- Refrigeration unit: low noise , fully hermetically sealed, air-cooled, low maintenance
- Casing parts made of stainless steel
- Insulated stainless steel hinges and hinged lid with magnetic seal
- Ram protection ring going over and under
Control ST71:
- Electronic temperature control with LED display
- Visual and acoustic alarm signal
Control ST100:
- Microprocessor-controlled adjustment using a membrane keyboard
- Clearly legible LED display
- Protection against inadvertent activation
Maximum reliability thanks to an alarm system that is separate from the mains, with battery capacity for approx. 72 hours
- Display of the maximum temperature reached in the interior area during the fault
- Visual and acoustic alarm signal
Interfaces
- Potential-free contact with the connection to an external alarm system
- RS485 interface
- 100 mV/K output for connecting a temperature recording system