Edwards Vacuum A73106983 XDS46iC Small Dry Scroll Pump, 100-120/200-230V 1 ph 50/60 Hz set to 230V

 

Technical Data

Displacement 50 or 60Hz 60 m3h-1 / 35 ft3min-1
Peak pumping speed 50 or 60Hz 40 m3h-1 / 23.5 ft3min-1
Ultimate vacuum (Total pressure) 5 x 10-2 mbar / 4 x 10-2 Torr
Max continuous inlet pressure 40 mbar / 30 Torr
Max continuous outlet pressure 0.2 bar gauge / 2.9 psig
Max allowed inlet and GB pressure 0.5 bar gauge / 7 psig
Motor power 0.52kW
Power connector IEC EN60320 C19
Nominal rotation speed 1750 rpm
Weight 48 kg / 105 lb
Inlet flange NW40
Exhaust flange NW25
Noise level 55.4 dB (A)
Vibration <4.5mms -1 (rms)
Leak tightness (static) <1 x 10-6 mbar ls-1
Operating temperature range 10 to 40 °C / 50 to 104 °F
  • £9910.00
Qty:  
XDS scroll pumps have become industry standard when dry pumping is essential, proving to be a robust and clean vacuum pump solution in a range of applications and processes.

The XDS35i pump has an innovative bearing shield that isolates the vacuum environment from all forms of lubricant, making it not only lubricant-free but hermetically sealed. XDS35iNGB variant has had the gas ballast feature removed for applications such as rare gas recirculation and gas recovery.

The XDS46i shares many of the same features of the XDS35i but with a peak speed of 40 m3h-1. The pump has been optimised for maximum pumping speed at inlet pressures between 1 mbar and 10 mbar, making it ideally suited for LCMS and ICPMS applications requiring higher pumping speeds.

The C variant has been modified so that it is more suitable for use on vapour handling processes and may be used in some applications using corrosive substances. The C version is fitted with Chemraz® internal valve pads and stainless steel exhaust port.


Chemraz® is a registered trademark of Greene Tweed

Technology

Scroll pumps use intermeshing fixed and orbital scrolls to form non-contacting, crescent shaped gas pockets that are continuously compressed by the orbiting action. This means that, unlike oil-sealed rotary pumps, there is no sealant or lubricant required in the swept volume of the pump.

Applications

• Automotive; brake line and air conditioning evacuation
• Backing turbomolecular  pumps
• Chemical applications including gel dryers and solvent recovery
• Coating systems
• High energy physics
• Mass spectrometry
 
 
    • SKU
      A73106983
  • Manufacturer
    Edwards Vacuum